Previous: Energy Resolution Up: Energy Resolution Next: Energy Resolution

Uniformity of the Calorimeter Response

Non-uniformities in the response of the calorimeter affect its performance if they cannot be corrected by calibration. In the case of a liquid krypton calorimeter with a tower readout structure the electromagnetic shower profile is rather narrow and very steep near its axis. The average shower profile as measured with 120 GeV electrons is shown in figure . Up to 50 % of the visible shower energy is deposited in the central calorimeter cell.

There are two main sources for a non-uniformity of the calorimeter response with tower readout. The first is the mechanical tolerance of the cell width and the second is given by the presence of the charge collecting electrodes. These electrodes are made of passive material and only part of the charge produced close to the electrodes is seen due to the effective pulse shaping time of about 60 ns. Both effects would average out in case of uniform ionization across the calorimeter. But, this is not the case for a shower profile as shown in figure . Both a left-right asymmetry in the two half-cells forming one calorimeter cell and a decrease of the charge close to the readout electrodes were observed in the data. In the following, both effects are discussed in more detail.

For cells with an uniform electric drift field the initial current is given by

where is the total ionization charge deposited in the cell, the drift velocity and the width of a cell. In the case of the showers perpendicular to the direction of drift and a cell size of the order of half a Moliere radius, the density of the ionization is changing strongly over the width of a cell. Then the precision with which the cell width is known can become a limiting factor in the resolution of the calorimeter. Since each cell consists of two half-cells, deviations from the nominal width tend to be opposite for adjacent half-cells and left-right asymmetries in the two half-cells can be observed.

Figure shows the response of the calorimeter in one cell as a function of x, the coordinate of the electron impact point perpendicular to the readout electrodes, determined using the drift chamber information. This cell was chosen since it showed the largest observed left-right asymmetry.

During the last period of the test run, extra spacers were inserted to improve the constancy of the foil spacing. The improvement of the response uniformity can be seen in figure .

Monte Carlo simulation [12] confirmed that in order to maintain the energy independent term in the resolution to less than 0.5 %, the mechanical tolerances defining the width of the cells must be kept to better than 1 % (i.e. 100 m) for the geometry of this prototype.

The second effect is related to the fraction of ionization deposited in the passive material of the electrodes and to the effective shaping time during which the initial current is measured. Since is of the order of 60 ns, the ionization deposited in a region distant from the anode by less than does not fully contribute to the initial current measurement. As a consequence, the current loss is small if the shower core is far away from the anode, but a larger amount of charge is lost if the particle impinges close to the anode. Independently of integration time, a larger fraction of ionization ends in passive material when the shower is closer to either anode or cathode. As a result, the response of the calorimeter as a function of the impact position shows depletions at the positions of the electrodes. These depletions are less pronounced at the cathodes.

To minimize non-uniformities, a small angle zig-zag of the readout electrodes is planned for the final calorimeter. In order to study the benefits of a small angle between the calorimeter cell structure and the beam axis, data were taken at calorimeter orientations of 0, 13 and 26 mrad with respect to the beam axis. The depletion present at the centre of the cell is reduced by this rotation. Typical values for the depletion are reported in table . The tilt smeared out and hence decreased the non-uniformities.

The remaining non-uniformities have been measured and subsequently corrected offline for the determination of the final energy resolution.

PC