File : cal_WA-X5B_9708.txt ------------------------------------------------------ Testbeam August, 1997 in WEST Area, X5B, 100GeV/c pions from : August, 20. 97 to : August, 27. 97 ------------------------------------------------------ DBDS-71 on Scirocco 5 : (30.5+-2) e/ADC (TP,970825) (12.2+-1) e/ADC (TP,970825) DBDS-67 on Scirocco 6 : (12.5+-1) e/ADC (TP,970825) DBDS-47 on Scirocco 7 : (11.5+-1) e/ADC (TP,970825) People present during calibration Harris, Rudi, Dirk The calibration value of DBDS-71 in Scirocco 5 changed during calibration. Lab measurements show that the calibration (e/mV) of all three repeator boards is approximately the same (shown with TP). The calibration on DBDS-71 was originally 30.5e/ADC. After plugging cables on the repeatorboard to the ADC back and forth we find the calibration changed to 12 e/ADC. dme/970907