Table Of The CSCTF Errors During Local and Global Runs
Day |
Time |
Error |
Action Taken |
Possible Reason |
Time Invested |
Type of Run |
elog |
|
|
|
|
|
|
|
|
04/06/09 |
18:37 |
Greg observes some strange behavior at the trigger and he wants to be sure we are in the correct state |
cleaned up RCMS from previous local run(1), restarted the csctf-cell and configured. Checked the localCondition.log that all SPs are properly configured |
This morning I was working on CSC-DT Synchronization and I played with the configuration. I did not clean up before leaving P5 |
~ 5 mins |
Local Run |
|
08/06/09 |
18:15 |
Philip, csc shifter, observes the TF DDU counting a huge amount of events; Karoly called me since he wants to be sure we are in the correct stateand the TF side is clean |
Checked the localCondition.log that all SPs are properly configured and checked the rate from the rate monitor on the cell; everything seems fine from the TF side |
|
~ 5 mins |
Local Run |
elog |
17/06/09 |
07:00 |
TF DDU reports error on local DAQ monitor |
TF DDU does not reports signal from SP2 (orange led not blinking at TF DDU input 1/B); I swapped fiber link between SP1 and SP2 and the error moved to SP1 input. Clearly it is a fiber error. Replaced the fiber. |
input 1/B of fiber link from SP2 to TF DDU broken |
~ 2 hours (most of the time wasted in commuting from P5 to 904) |
Local Run |
elog 64361 |
23/06/09 |
10:15 |
Major leakage on the plus endcap; it is turned off: need to mask the FMM errors |
Plus endcap masked; restarted the csctf-cell; configured and checked the configuration |
Leakage on the plus endcap |
10 mins |
Global Run |
elog 65526 |
(1) destroy local, startTF, DQM, DQM_Display and DAQ_WriteY_BuildY
Counters of Errors
Error |
Configuration Check |
Hardware |
Crate Masking |
# Times |
2 |
1 |
1 |
--
GianPieroDiGiovanni - 05 Jun 2009