Medipix2

Features

  • High spatial, high contrast CMOS pixel readout chip
  • Single photon counting mode (noise suppresion, superior SNR)
  • Direct X-ray conversion
  • accepts +/- charge
  • Detector leakage current compensated pixelwise
  • window in energy selectable
  • choosable exposure times, data accumulated in 13bit counter/pixel
  • Parallel/serial readout after exposure
  • 2 threshold levels, allowing multi-spectral medical imaging
  • noise per channel: ~110e- RMS

Parameters

  • pixel size: 55μm
  • CMOS technology: 0.25μm
  • count rate per pixel: 300KHz
  • active area: 2cm^2
  • 256x256 pixel matrix

Applications

  • X-Ray Imaging
    • X-Ray Diffraction (due to chip's high dynamic range) - commercialized by PANalytical
    • Phase contrast imaging with microfocus X-ray tubes
    • CT and Micro-CT
    • used as prototype digital readout for mammography, dental imaging, multi-spectral CT imaging
  • Electron microscopy
    • electron cryo-microscopy
    • Low Energy Electron Microscopy (LEEM) (benefits from low noise and high dynamic range)
  • Visible Single Photon Imaging
  • Neutron Monitoring
  • Tracking and Vertexing
  • Mass Spectrometry
    • miniaturisation of molecular imaging mass spectrometry systems
  • Sensor development

-- SotirisFragkiskos - 05-Sep-2011

Edit | Attach | Watch | Print version | History: r4 < r3 < r2 < r1 | Backlinks | Raw View | WYSIWYG | More topic actions
Topic revision: r4 - 2011-09-07 - SotirisFragkiskos
 
    • Cern Search Icon Cern Search
    • TWiki Search Icon TWiki Search
    • Google Search Icon Google Search

    Main All webs login

This site is powered by the TWiki collaboration platform Powered by PerlCopyright &© 2008-2024 by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
or Ideas, requests, problems regarding TWiki? use Discourse or Send feedback